Found 2 results

510(k) Data Aggregation

    K Number
    K192745

    Validate with FDA (Live)

    Date Cleared
    2019-11-26

    (57 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DePuy Synthes Locking Condylar Plate (LCP) System is intended for buttressing multifragmentary distal femur fractures including: supracondylar, intra-articular and extraarticular condylar fractures, fractures in normal or osteopenic bone, and non-unions and malunions.

    The DePuy Synthes Sacral Bar System is intended for fixation of fractures of the posterior pelvis, in areas of the posterior superior iliac spine and posterior iliac spine, for sacral fractures and fracture-dislocations of the sacro-iliac joint.

    The DePuy Synthes Spiked Washer is intended for use in ligament reattachment or fixation, specifically re-adaptation of torn or avulsed ligaments.

    The DePuy Synthes 4.0 mm Titanium Locking Screws are intended to be used with existing DePuy Synthes LCP® plating systems for the fixation of various long bones, such as the humerus, femur and tibia.

    The DePuy Synthes 5.0/7.3 mm Cannulated Locking Screws are intended to be used with existing DePuy Synthes LCP® plating systems for the fixation of various long bones, such as the femur.

    The DePuy Synthes Peri-Prosthetic Screws are intended for fixation of various long bones, such as the humerus, femur and tibia, in conjunction with DePuy Synthes Locking Plates that accept 4.0/5.0 mm locking screws. They are also for use in fixation of periprosthetic fractures, osteopenic bone, and fixation of nonunions or malunions.

    The DePuy Synthes 3.5 mm Conical Screws are intended to be used with existing DePuy Synthes LCP plating systems for fixation and interfragmentary compression of various bones, such as the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, tibia and fibula, particularly in osteopenic bone.

    The DePuy Synthes Spherical Washers are intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/load over a large area when used for fracture fixation of large (long) bone and bone fragments. In addition, the DePuy Synthes Spherical Washers are intended to prevent the projection of the screw head, when the screw must be inserted at an acute angle (e.g., in ankle arthrodesis).

    DePuy Synthes 6.5 mm Cancellous Screws are intended for use in hindfoot and midfoot fusions, subtalar fusions, calcaneal osteotomies, midfoot reconstruction, and ankle arthrodeses.

    DePuy Synthes 6.5 mm Midfoot Fusion Bolt is indicated for fracture fixation, osteotomies, nonunions, and fusions of large bones in the foot and ankle.

    The DePuy Synthes 1.0mm, 1.5mm, 2.0mm, and 2.4mm Cortex Screws are intended for use in trauma procedures, reconstructive procedures, and general surgery of the hand, wrist, and other small bones and bone fragments in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. The DePuy Synthes 2. 7mm Cortex Screw is intended for fractures and osteotomies of small bones and bone fragments, including the foot, ankle, and hand in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. The DePuy Synthes 2. 7mm Cortex Screw may also be used in fusion applications in adults and adolescents (12-21 years) when used with the DePuy Synthes 2.4mm/2.7mm Variable Angle LCP Forefoot/Midfoot System (K100776) and in adults and pediatric patients (2-12 years) when used with the DePuy Synthes Ti Wrist Fusion Plate (K023879). The DePuy Synthes 3.5mm and 4.0mm Cortex Screws are intended for fixation of fractures, osteotomies and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, tibia, calcaneous, femur and fibula in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. The DePuy Synthes 4.5mm Cortex Screw is intended for fixation of various long bones, such as the humerus, femur and tibia. It is also for use in fixation of non-unions or mal unions in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation.

    The DePuy Synthes 4.0 mm Cortex Screws are intended for fixation of fractures, fusions, osteotomies, non-unions, and malunions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, tibia, calcaneus, femur, fibula and the bones of the foot and hand in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. The DePuy Synthes 4.5 mm Cortex Screws are intended for fixation of fractures, fusion, osteotomies, non-unions, and malunions of various long bones, such as the humerus, femur and tibia; the pelvis of the foot and hand in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. The DePuy Synthes 2.4 mm Cannulated Screws are intended for fixation of fractures, fusions, osteotomies, non-unions, and malunions of small bones in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. The DePuy Synthes 3.5 mm and 4.0 mm Cannulated Screws are intended for fixation of fractures, fusions, osteotomies, non-unions, and malunions of small bone fragments in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. The DePuy Synthes 4.5 mm Cannulated Screws are intended for fixation of fractures, fusions, osteotomies, non-unions, and malunions of long bone fragments; and the bones of the hand and foot, in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. The DePuy Synthes 6.5 mm Cannulated Screws are intended for fixation of fractures, fusions, osteotomies, non-unions, and malunions of long bones and long bone fragments, pelvis, sacrum, and the bones of the foot in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. These screws are also indicated for femoral neck fractures; slipped capital femoral epiphysis; an adjunct to DHS in basilar neck fractures; tibial plateau fractures; ankle arthrodesis; pediatric femoral neck fractures; intercondylar femur fractures; SI joint disruptions; and subtalar arthrodesis. The DePuy Synthes 7.0 mm and 7.3 mm Cannulated Screws are intended for fixation of fractures, osteotomies, non-unions, and malunions of long bone fragments, pelvis, sacrum, and the bones of the foot in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. These screws are also indicated for slipped capital femoral epiphysis; ankle arthrodesis; and subtalar arthrodesis. The DePuy Synthes 1.5 mm Headless Compression Screws are intended for fixation of intraarticular and extra-articular fractures, avulsions, malunions, and osteotomies of small bones and small bone fragments, as well as arthrodesis of small joints in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. The DePuy Synthes 2.4 mm Headless Compression Screws are intended for fixation of fractures, osteotomies, non-unions, and malunions of small bone arthrodesis in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. The DePuy Synthes 3.0 mm Headless Compression Screws are intended for fixation of intraarticular and extra-articular fractures, non-unions, and osteotomies of small bones and small bone fragments, as well as arthrodesis of small joints in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. The DePuy Synthes 4.5 mm and 6.5 mm Headless Compression Screws are indicated for fixation of fractures, fusions, osteotomies, non-unions, and malunions of various bones and bone fragments including bones of the foot, humerus, femur and tibia in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation.

    The DePuy Synthes 2.0 mm Quick Insertion Screws are intended for fixation of fractures, fusions, osteotomies, nonunions, and malunions of the forefoot, midfoot, and hand in patients 2 years of age and older in which growth plates have fused or in which growth plates will not be crossed by screw fixation.

    The DePuy Synthes ALPS is a plate and screw system intended to treat fractures of various long bones, including the radius, ulna, humerus, tibia, fibula, and femur.

    The DePuy Synthes Sterile 3.0 mm Cannulated Screw, used with the Threaded Washer, is generally intended for intra-articular fixation of small bones, such as the hand, wrist and forefoot. Specifically, it is intended for fractures of carpals and metacarpals, carpal and metacarpal arthrodesis; small fragments of the hand and wrist, and certain metatarsal phalangeal applications (in foot). The 3.0 mm Cannulated Screw, by itself is intended for fixation of small bones, such as the hand, wrist and forefoot.

    The DePuy Synthes CABP System is a plate and screw system intended to treat fractures of the proximal humerus and distal tibia. Proximal humerus fractures include two-part greater tubercle fractures and fracture dislocations, two-part surgical neck fractures and fracture dislocations, three-part fractures or fracture dislocations, fractures in osteopenic bone, and nonunions and malunions. Distal tibia fractures include acute fractures in osteopenic bone, and nonunions and malunions.

    Device Description

    The 4.0 mm locking screws feature a self-tapping tip, are solid, and have a flat head profile with rounded edges. They are available in lengths ranging from 14 mm to 62 mm. The threads just below the head of each locking screw engage with the threaded holes of the plate. The engagement of the screw to the plate shaft creates a locked fixed angle construct.

    The DePuy Synthes Sacral Bar System consists of a threaded bar, washers, and nuts. The bars are fully threaded. One end of the bar has a trocar point to guide the bar through pre-drilled holes. The bars are available in lengths ranging from 120 to 260 mm, in 10 min increments. The washers that are used with this system are oval shaped and are designed to slide freely along the bars. Both rounded and straight nuts are provided with this system; the rounded nuts mate with the washers to create compression, while the straight nuts are then added to wedge against the rounded nuts to maintain compression.

    The DePuy Synthes Spiked Washers are made of PEEK Optima™ LT plymer + 6% barium sulfate. There are either six or eight spikes arrayed around a center hole that will accommodate either 2.7, 3.5, 4.0, 4.5 or 6.6 mm screws.

    The DePuy Synthes 4.0 mm Titanium (TI) Locking Screws feature a self-tapping tip, stardrive mechanishm, and have a flat head profile with rounded edges. They are available in lengths ranging from 14 mm to 90 mm. The threads below the head of each locking screw are designed to engage with the threaded holes of currently marketed DePuy Synthes LCP plating systems.

    The DePuy Synthes 5.0/7.3 mm Cannulated Locking Screws feature self-drilling and self-tapping tips, have a flat head profile with rounded edges with a hex drive recess. They are available in additional lengths ranging from 100 – 145 mm. The threads below the head of each locking screw are designed to engage with the threaded holes of currently marketed DePuy Synthes LCP® plating systems

    The DePuy Synthes Peri-Prosthetic Screws feature a self-tapping blunt tip, stardrive mechanism, and have a flat head profile with rounded edges. They are available in lengths ranging from 8 mm to 12 mm. The threads on the head of each locking screw are designed to engage with the threaded holes of currently marketed DePuy Synthes LCP® plating systems.

    The DePuy Synthes 3.5 mm Conical Screws are self-tapping, have a stardrive or hex drive recess, and are manufactured from stainless steel and titanium. The 3.5 mm conical screws are available in various lengths ranging from 40-95 mm.

    The DePuy Synthes Spherical Washers are round or oval in design with a slotted center hole which allows for screw angulation up to 70 degrees. They are used with 4.5 - 7.3 mm diameter screws and manufactured from Stainless Steel and Titanium.

    The DePuy Synthes 6.5 mm Cancellous Screws incorporate a fully threaded shaft, 4.0 mm core diameter, and have a flat head profile with rounded edges. They are available in lengths ranging from 60 mm to 130 mm in both Stainless Steel and Titanium Alloy. The screws are provided STERILE and NON STERILE.

    The DePuy Synthes 6.5 mm Midfoot Fusion Bolt is a solid metallic fixation bolt which is intended to be used in procedures involving the foot and ankle. The DePuy Synthes 6.5 mm Midfoot Fusion Bolt is partially threaded on both ends and will be available in versions composed of implant quality stainless steel and titanium alloy.

    The DePuy Synthes Cortical Screws have self-tapping features, stardrive, or cruciform head recesses, and are manufactured from stainless steel, commercially pure titanium, and/or titanium alloy. Cortex screws are offered both sterile and non sterile and are available in various lengths. Screws may be used independently or with any DePuy Synthes plate which accepts 1.0 mm, 1.3 mm, 2.0 mm, 2.4 mm, 2.7 mm, 4.0 mm, and 4.5 mm cortex screws. The screws, when used in pediatric applications, may be used independently or with compatible Synthes plates which are also indicated for pediatric populations.

    The DePuy Synthes Cortex, Cannulated, and Headless Compression Screws are metallic bone screws manufactured from Stainless Steel (ASTM F138), Commercially Pure Titanium (ASTM F67), and/or Titanium Alloy (ASTM F1295). The screws are available in multiple lengths and diameters, and are intended to be used as stand-alone bone screws for internal bone fixation of fractures, fusions, osteotomies, non-unions, and malunions in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by the screw fixation. The DePuy Synthes Cannulated and Headless Compression screws are cannulated for use with guidewires for precise placement in bone with the exception of the 1.5 mm Headless Compression Screw, which is a solid screw and therefore does not allow for instrumentation with a guide wire. The Headless Compression Screws feature threaded heads that allow for purchase in the near cortex of bone during and after implantation, potentially reducing complications associated with countersinking of traditional cortex or cannulated screws.

    The DePuy Synthes 2.0 mm Quick Insertion Screws (QI Screws) are twist-off style screws comprised of a screw and post which are intended to separate at a predefined location (the breakpoint) as the screw is implanted into bone. The screws are offered with an external thread diameter of 2.0 mm and in lengths ranging from 11 mm to 18 mm to accommodate varying patient anatomy and fracture patterns in the foot and hand. The screws are made of a titanium alloy (Ti-6Al-7Nb) in accordance with ASTM F1295.

    The DePuy Synthes ALPS is a plate and screw system intended to treat fractures of various long bones, including (but not limited to) the radius, ulna, humerus, tibia, fibula and femur. The ALPS utilizes a locking feature that secures the screw to the plate, enabling stable fixation to be achieved via unicortical or bicortical fixation. The locking feature consists of conical shaped, threaded screw holes in the plate that correspond wi th the conical-shaped, threaded head of the screw.

    The 3.0 mm Cannulated Screw has a 3.0 mm thread diameter and a 4.0 mm head diameter. It is available in short and long thread lengths ranging from 10 mm to 40 mm. It features selfdrillinglself-tapping flutes and cancellous threads for use in cancellous bone. The screw is used in conjunction with a guide wire for precise placement in bone. The Threaded Washer has a 5.5 mm thread diameter and is 3.0 mm in length. It features exterior threads for bone purcbase in the near cortex and is cylindrically shaped to allow passage of the 3.0 mm Cannulated Screw. The washer serves to provide a buttress in the near cortex for which the screw head can compress against.

    The DePuy Synthes Cannulated Angle Blade Plate is a straight plate with a blade at the head to allow for better fixation in the head of the humerus or in the distal tibia. The blade of the plate is cannulated to fit over a guide wire, allowing for the adjustment of the wire placement several times without adversely affecting the final result. There are cuts in the undersurface of the plate to reduce the surface area of the plate in contact with bone. The plate accepts 4.5 mm and 6.5 mm screws. he Cannulated Angle Blade Plate System is manufactured from stainless steel or commercially pure titanium.

    AI/ML Overview

    This document describes the FDA's clearance of several DePuy Synthes trauma screws and plates as "MR Conditional" devices. The clearance is based on non-clinical testing demonstrating their safety in an MR environment, not on clinical performance or diagnostic accuracy. Therefore, the questions related to clinical study design (sample size, ground truth, expert adjudication, MRMC studies, standalone performance) are not applicable in this context.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    Acceptance Criteria (from ASTM standards)Reported Device Performance (Summary)
    Magnetically Induced Displacement Force (ASTM F2052-14):Testing demonstrated conditional safety with respect to magnetically induced displacement force.
    Magnetically Induced Torque (ASTM F2213-06):Testing demonstrated conditional safety with respect to magnetically induced torque.
    Radiofrequency (RF) Heating (ASTM F2182-11a):Testing demonstrated conditional safety with respect to RF heating.
    Image Artifacts (ASTM F2119-07):Testing demonstrated conditional safety with respect to image artifacts.
    Overall MR Safety (MR Conditional Labeling):All listed devices (DePuy Synthes Locking Condylar Plate (LCP) System, Sacral Bar System, Spiked Washer, 4.0 mm Titanium (TI) Locking Screws, 5.0/7.3 mm Cannulated Locking Screws, Peri-prosthetic Screws, 3.5 mm Conical Screws, Spherical Washers, 6.5 mm Cancellous Screws, 6.5 mm Midfoot Fusion Bolt, Cortical Screws, Cortex Screws, Cannulated Screws, Headless Compression Screws, 2.0 mm Quick Insertion Screws, Anatomical Locking Plate System (ALPS), Sterile 3.0 mm Cannulated Screw and Threaded Washer, Cannulated Angle Blade Plate (CAEP) System) meet the criteria for MR Conditional labeling.

    2. Sample size used for the test set and the data provenance

    Not applicable. The study was non-clinical (in-vitro) testing of the devices' physical properties in an MR environment, not a test set using human data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The "ground truth" here is adherence to specific ASTM standards for MR safety, which are technical measurements rather than expert human interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This was non-clinical testing against engineering standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This submission is for the MR Conditional status of medical implants, not for an AI diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This submission is for the MR Conditional status of medical implants, not for an AI diagnostic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used was compliance with established ASTM (American Society for Testing and Materials) standards for evaluating the MR safety of medical implants. These standards involve specific experimental setups and measurement methodologies to assess:

    • Magnetically induced displacement force (ASTM F2052-14)
    • Magnetically induced torque (ASTM F2213-06)
    • Radiofrequency (RF) heating (ASTM F2182-11a)
    • Image artifacts (ASTM F2119-07)

    8. The sample size for the training set

    Not applicable. This was non-clinical testing of physical devices. No training set for an algorithm was involved.

    9. How the ground truth for the training set was established

    Not applicable. No training set for an algorithm was involved.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181933

    Validate with FDA (Live)

    Date Cleared
    2019-05-09

    (294 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DePuy Synthes Sterilization Container System is a device intended to be used to enclose other medical devices to be sterilized by healthcare facilities. It allows sterilization of the enclosed medical devices and maintains sterility of the devices until used for a maximum of 180 days.

    The containers are suitable for dynamic air removal (pre-vacuum) steam sterilization when used according to the instructions for use.

    Device Description

    The DePuy Synthes Reusable Sterilization Container System is a rigid, reusable, container intended to be used to sterilize other medical devices and maintain sterility of these devices until used. The container system is comprised of a lid with gasket, base, filter, tamper evident arrows, and data cards. The container system includes a lifting platform to hold Synthes graphic cases containing orthopedic medical devices (instruments and implants) within the container.

    AI/ML Overview

    The DePuy Synthes Sterilization Container System (K181933) is intended to enclose medical devices for sterilization and maintain their sterility for up to 180 days. The device uses dynamic air removal (pre-vacuum) steam sterilization. No clinical tests were performed, but extensive non-clinical tests were conducted to demonstrate substantial equivalence to the predicate device, Synthes Reusable Sterilization Container System (K130720).

    1. Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Sterilization Efficacy: Pre-Vacuum SteamObjective: To validate the efficacy of the DePuy Synthes Sterilization Container System using pre-vacuum sterilization cycle of 270°F (132°C) with 4 minutes of exposure.
    Result:Testing demonstrated a 12 log reduction and a sterility assurance level (SAL) of 10⁻⁶ using the biological (BI) overkill method and half-cycle validation.
    Dry TimeObjective: To determine the drying time for the DePuy Synthes Sterilization Container System following a steam pre-vacuum cycle of 270°F (132°C) with 4 minutes of exposure.
    Result:Dry time studies establish minimum dry time of 30 minutes for pre-vacuum steam sterilization modality.
    180 Day Event Related Shelf LifeObjective: To validate the 180-day event related shelf life evaluation of the DePuy Synthes Sterilization Container System following exposure to a full pre-vacuum sterilization cycle (lethality plus dry time).
    Result:180 Day Event Related Shelf life studies demonstrated sterility maintenance for the recommended pre-vacuum steam sterilization modality. The device maintains sterility for a maximum of 180 days.
    Microbial ChallengeObjective: To demonstrate the DePuy Synthes Sterilization Container System can maintain package integrity/microbial barrier after being subjected to sterilization processing and a whole package microbial (aerosol) challenge.
    Result:Whole package microbial challenge test, exposing a container to a minimum of 1 x 10⁶ Bacillus atrophaeus colony forming units (CFU) via an aerosol challenge demonstrating 100% negative growth.
    Limits of ReuseObjective: To verify the limits of reuse of the DePuy Synthes Sterilization Container System when processed in repetitive pre-cleanings, mechanical washer cycles and steam pre-vacuum cycles.
    Result:After 100 cycles of processing samples showed no signs of physical or mechanical degradation and continued to function as intended.
    BiocompatibilityObjective: To evaluate the biocompatibility of the DePuy Synthes Sterilization Container System for cytotoxicity to mammalian cells in culture by a method compliant with the requirements specified in ISO 10993-5:2009.
    Result:Results showed no heavy metal leaching or cytotoxicity at detectable levels. The device materials are toxicologically acceptable for their intended use. For specific materials, stainless steel, aluminum, silicone (gasket), SMS polypropylene (filter) were used in the construction and were demonstrated to withstand repeated processing with proven compatibility with the sterilization process and sterilants. Examples of intrinsically stable metals: stainless steel, titanium (CP and alloys), aluminum. Thermoplastic polymers: PEEK, PEI (Ultem), Acetal (Delrin), Radel (PPSU), Nylon, PTFE, Polypropylene, ABS (Acrylonitrile Butadiene Styrene), POM (Polyoxymethylene). Thermosetting polymers: Phenolic, Silicone. Composites: carbon fiber reinforced epoxy (CFRE). The materials are all compliant with the criteria for the stated materials.

    2. Sample size used for the test set and the data provenance

    The document indicates that testing was performed on "samples" of the device, container models, and worst-case configurations of joint reconstruction devices. Specific numerical sample sizes for each test are not explicitly provided. The data provenance is described as non-clinical testing performed by the manufacturer, DePuy Synthes. The studies are retrospective as they summarize verification and validation testing results.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This submission is for a sterilization container system, and the evaluation is based on engineering and microbiological performance studies (non-clinical testing). Therefore, the concept of "ground truth established by experts" in the context of interpretation (e.g., radiologist for imaging) is not directly applicable. The "truth" is established by adherence to recognized standards (e.g., ANSI/AAMI ST77:2006, ANSI/AAMI ST79:2010, ISO 10993-5:2009, USP General Chapter <231> Heavy Metals) and objective measurements performed by qualified personnel in laboratories following Good Laboratory Practices (GLP), implied by the regulatory context.

    4. Adjudication method for the test set

    Not applicable, as this is a non-clinical device performance study based on objective measurements against scientific standards, not a diagnostic or interpretative study requiring expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical device sterilization container, not an AI-assisted diagnostic or imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a medical device sterilization container, not an AI-based algorithm.

    7. The type of ground truth used

    The ground truth for the non-clinical tests is based on:

    • Microbiological Standards: Demonstrated sterility assurance level (SAL) of 10⁻⁶ and 12 log reduction via biological indicators (BIs) (for sterilization efficacy and microbial challenge).
    • Physical and Chemical Measurement Standards: Adherence to established standards for dry time, material properties (e.g., heavy metals leaching, cytotoxicity), and physical integrity (e.g., limits of reuse, package integrity).
    • Industry and Regulatory Standards: Compliance with ANSI/AAMI ST77:2006, ANSI/AAMI ST79:2010, ISO 10993-5:2009, and USP General Chapter <231> Heavy Metals.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable.